This article covers control file setup for panels that are more complex than a straightforward single-cell-type staining with one unstained sample. Specifically:

  1. Mixed bead and cell controls — different universal negatives for each type
  2. Viability markers — how the is.viability column affects the legacy gating pipeline (not the automated pipeline)
  3. Fluorophores not in the AutoSpectral database — how to handle them and how to fill in their peak channel manually
  4. Fluorophore name disambiguation when you have both beads and cells for the same target (not recommended, but possible)

This builds on the basics in 02 Creating the Control File. The automated pipeline (get.spectra.automated()) is used throughout.

library(AutoSpectral)
asp <- get.autospectral.param(cytometer = "aurora")
control.dir <- "~/my_experiment/controls"
create.control.file(control.dir, asp)
control.file <- "fcs_control_file.csv"

Mixed bead and cell controls

Many panels use compensation beads for dim or rare markers and cell controls for bright or common ones. Both types can coexist in a single control file.

The control.type column tells AutoSpectral how to treat each sample. Set each row to "cells" or "beads" (lower case). In the automated pipeline:

  • Cells are processed with AF orthogonalisation and cosine-similarity filtering against the universal negative’s AF vector. Scatter-matched AF subtraction uses the nearest events from the paired unstained cell sample.
  • Beads are assumed to have homogeneous, low-AF backgrounds. Cosine filtering against an AF vector is skipped. The “universal” negative for a bead control should be an unstained bead sample. If your beads have separate negative and positive beads, use the unstained positive beads as the universal negative.

check.control.file() enforces that the control.type of a sample and its universal.negative must match — a cell control paired with a bead negative (or vice versa) is flagged as an error.

Naming fluorophores when you have both beads and cells

check.control.file() requires every fluorophore value to be unique. If you have both a bead and a cell single-stained control for the same fluorophore, give them distinct names. For example:

filename fluorophore control.type universal.negative
PE_cells.fcs PE cells Unstained_cells.fcs
PE_beads.fcs PE beads beads Unstained_beads.fcs
Unstained_cells.fcs AF cells Unstained_cells.fcs
Unstained_beads.fcs Negative beads Unstained_beads.fcs

The name in the fluorophore column becomes the parameter name in the unmixed FCS file, so pick whichever version you want to appear in your analysis software. You will then run get.spectra.automated() once and receive spectra for both PE and PE beads. Choose one for unmixing — typically whichever produces the cleaner, more accurate spectrum (check the QC ribbon plots in figure_spectral_ribbon/).

If you want to unmix using only one, subset the spectra matrix before passing it to unmix.fcs():

# keep only the cell-based PE spectrum
spectra.subset <- spectra[ rownames(spectra) != "PE beads", ]

Multiple unstained samples

When you have more than one unstained sample (one for cells, one for beads), both must appear in the filename column. Set fluorophore to AF for the unstained cell sample and to Negative for the unstained bead sample. check.control.file() enforces that every value in universal.negative must itself appear as a row in filename, and that it must be labelled AF or Negative.

issues <- check.control.file(control.dir, control.file, asp)

Viability markers

A viability dye labels dead cells selectively. Dead cells sit to the left of the live population on FSC. This only matters for the legacy pipeline (define.flow.control()), which uses the is.viability column to widen the FSC gate leftward to capture dead cells. The column is ignored by get.spectra.automated().

If you are using the automated pipeline, you do not need to fill in is.viability. The automated extraction selects high-expressing events based on the peak channel signal and cosine filtering rather than scatter gating, so it works correctly for viability dyes without any special configuration.

For completeness, the legacy approach is:

| filename           | fluorophore | is.viability |
|--------------------|-------------|--------------|
| LiveDead_cells.fcs | eFluor780   | TRUE         |

For the v1.5.0 landmark gating system, assign the viability control to a dedicated gate group by setting gate.name to something like "dead" and gate.define = TRUE. See the Full Workflow and Advanced Gating articles for details.

Fluorophores not in the database

If AutoSpectral cannot match a file name to a known fluorophore, you will see No Match in the fluorophore column. This is a validation error — check.control.file() will not let you proceed until every No Match is replaced.

Step 1 — fill in fluorophore. Write any name you like. This name will appear in the unmixed FCS file and in all AutoSpectral plots, so choose carefully. The name does not need to match any database entry for the pipeline to run.

Step 2 — fill in channel. This is the expected peak detection channel for the fluorophore on your cytometer. You can find it using:

In the automated pipeline, the channel column is optional — if you leave it blank, check.control.file() will issue a warning and get.spectra.automated() will derive the peak channel empirically from the data. However, providing the correct channel enables the QC step to verify the extraction against the expected peak, which is strongly recommended. An unrecognised fluorophore will not have a spectral reference library entry, so the cosine-similarity QC check against the reference library will be skipped for that fluorophore; only the peak-signal check will apply.

Step 3 (optional but encouraged) — add to the database. If you add your fluorophore (with synonyms and peak channel entries for each cytometer) to the fluorophore database, it will be recognised automatically by future versions of AutoSpectral. Changes are incorporated with the next package update.

Putting it together — example control file

Here is what a completed control file might look like for a panel with bead and cell controls, an unstained bead sample, a viability dye, and one fluorophore not in the database:

filename fluorophore marker channel control.type universal.negative
BV421_cells.fcs BV421 CD4 V4-A cells Unstained_cells.fcs
PE_beads.fcs PE CD14 YG3-A beads Unstained_beads.fcs
eFluor780_cells.fcs eFluor780 Viability cells Unstained_cells.fcs
NewDye660_cells.fcs NewDye660 CD8 R2-A cells Unstained_cells.fcs
Unstained_cells.fcs AF cells Unstained_cells.fcs
Unstained_beads.fcs Negative beads Unstained_beads.fcs

Notes on this example:

  • eFluor780 has no channel entry — the automated pipeline will derive the peak empirically and warn about the missing channel. In practice it is best to fill this in.
  • NewDye660 is a hypothetical fluorophore not in the database. Its channel (R2-A) has been filled in manually.
  • Both unstained samples are present, with matching control.type values for each paired control.

Running the check

issues <- check.control.file(control.dir, control.file, asp)

The check is run in automated mode (legacy = FALSE by default), so gating columns are not validated. Pay attention to:

Severity What to look for
Error No Match in fluorophore; duplicate fluorophore names; control.type mismatch between sample and its universal negative; universal negative filename not found; AF control absent or set to beads
Warning Missing channel for non-AF/Negative rows; event count < 5,000; cytometer name mismatch

Fix all errors before proceeding. When the check passes with a green message, run spectral extraction:

spectra <- get.spectra.automated(
  control.dir      = control.dir,
  control.def.file = control.file,
  asp              = asp
)

Check the outputs in figure_spectral_ribbon/ to verify that spectra for both bead and cell controls look reasonable, then select the best version of each for unmixing.